若您需要咨询产品或有任何技术问题,请通过官方电话 400 885 9050 或邮箱 info.cn@stemcell.com 与我们联系

ThawSTAR® CFT2测试样品管

用于每次解冻之前和之后记录仪器性能和审计跟踪

产品号 #(选择产品)

产品号 #100-0643_C

用于每次解冻之前和之后记录仪器性能和审计跟踪

总览

使用ThawSTAR® CFT2测试样品管记录ThawSTAR® CFT2全自动细胞解冻系统的性能。在与待解冻生物样本相同的条件下使用 ThawSTAR® CFT2测试样品管,评估并记录每次解冻前后的解冻性能,并创建审计追踪。为了方便起见,ThawSTAR® CFT2测试样品管可单独购买,也可以作为ThawSTAR® CFT2全自动细胞解冻系统的一部分购买。ThawSTAR® CFT2全自动细胞解冻系统配有两个ThawSTAR® CFT2测试样品管,ThawSTAR® CFT2转运盒,ThawSTAR® CFT2电源和适配器,可实现无水标准化细胞解冻流程。如需获取性能鉴定文件和功能测试所需的配件,请使用ThawSTAR® CFT2 IOPQ试剂盒(可单独购买)。

有关使用ThawSTAR®进行细胞解冻方案的更多信息,请浏览技术手册.

应用
冻存
 
品牌
ThawSTAR
 
研究领域
脐带血库,免疫
 

实验数据

Bar graphs comparing the number of live cells recovered and percentage viability for peripheral blood mononuclear cells (PBMCs) thawed using ThawSTAR® CFT2 Automated Thawing System versus a water bath.

Figure 1. Frozen PBMCs Thawed Using the ThawSTAR® CFT2 Automated Thawing System Show High Recovery and Viability

Peripheral blood mononuclear cells (PBMCs) cryopreserved in CryoStor® CS10 (Catalog #07930) at a concentration of 3 x 107 cells/vial were retrieved from liquid nitrogen, one week after storage. When thawed using the ThawSTAR® CFT2 Automated Thawing System or a water bath, (A) the mean live cell recovery was 2.78 x 107 vs. 2.96 x 107 cells, respectively and (B) the mean viability was 92.8% vs. 93.9%, respectively. The cells were from 3 different donors and were tested in triplicates. Cell recovery and viability was assessed using a Nucleoview counter.

Bar graphs comparing the number of live cells recovered and percentage viability for monocytes thawed using ThawSTAR® CFT2 Automated Thawing System versus a water bath.

Figure 2. Frozen Monocytes Thawed Using the ThawSTAR® CFT2 Automated Thawing System Show High Recovery and Viability

Monocytes cryopreserved in CryoStor® CS10 (Catalog #07930) at a concentration of 2 x 107 cells/vial were retrieved from liquid nitrogen, one week after storage. When thawed using the ThawSTAR® CFT2 Automated Thawing System or a water bath, (A) the mean live cell recovery was 2.05 x 107 vs. 1.89 x 107 cells, respectively and (B) the mean viability was 95% vs. 94.6%, respectively. The cells were from 3 different donors and were tested in triplicates. Cell recovery and viability was assessed using a Nucleoview counter.

Bar graphs comparing the number of live cells recovered and percentage viability for human pluripotent stem cells (hPSCs) thawed using ThawSTAR® CFT2 Automated Thawing System versus a water bath.

Figure 3. Frozen hPSCs Thawed Using the ThawSTAR® CFT2 Automated Thawing System Show High Recovery and Viability

Human pluripotent stem cells (hPSCs) cryopreserved in CryoStor® CS10 (Catalog #07930) at a concentration of 1 x 106 cells/vial were retrieved from liquid nitrogen, one week after storage. When thawed using the ThawSTAR® CFT2 Automated Thawing System or a water bath, (A) the mean live cell recovery was 9.05 x 105 vs. 9.35 x 105 cells, respectively and (B) the mean viability was 83.04% vs. 82.93%, respectively. The hPSCs were from 3 different cell lines (M001, 1C, and H9) and were tested in triplicates. Cell recovery and viability was assessed using a Nucleoview counter.

产品说明书及文档

请在《产品说明书》中查找相关支持信息和使用说明,或浏览下方更多实验方案。

Document Type
Product Name
Catalog #
Lot #
Language
Document Type
Safety Data Sheet
Catalog #
100-0643
Lot #
All
Language
English

应用领域

本产品专为以下研究领域设计,适用于工作流程中的高亮阶段。探索这些工作流程,了解更多我们为各研究领域提供的其他配套产品。

质量保证:

产品仅供研究使用,不用于针对人或动物的诊断或治疗。 欲获悉更多关于STEMCELL的质控信息,请访问 STEMCELL.CN/COMPLIANCE.
Copyright © 2025 by STEMCELL Technologies. All rights reserved.